TD4520 Portable Three-phase and DC Calibrator

TD4520 is a multi-function 3PH power calibrator. Class 0.02 / class 0.05 available. It integrates AC/DC power output function, 3PH measuring function, harmonic function, transmitter test function, U4th voltage output function, AC energy meter test function. Can be used to calibrate AC/DC voltmeter, AC/DC ammeter, AC/DC power meter electric energy meter and transmitter, etc.

1. Summary

TD4520 is a multi-function 3PH power calibrator. Class 0.02 / class 0.05 available. It integrates AC/DC power output function, 3PH measuring function, harmonic function, transmitter test function, U4th voltage output function, AC energy meter test function. Can be used to calibrate AC/DC voltmeter, AC/DC ammeter, AC/DC power meter electric energy meter and transmitter, etc.

2. Features

• 3PH voltage output:6 V~456 V

• 3PH current output:20 mA~24 A

• 3PH voltage meter:6 V~456 V;

• 3PH current meter:0.1 A~6 A

• Frequency:45 Hz~70 Hz

• U-I phase:0°~360°

• DC voltage output:10 mV~660 V

• DC current output:0.1 mA~22 A

• Accuracy class: 0.02, 0.05.

• 2nd~21st harmonic output

• Transmitter test function

• U4th AC voltage output function(option)

• Energy meter test function(option)

• Test software(option)

3. Application

 

 

Image 1

It can meet most single-phase / three-phase meters or DC meters.
Image 2

• It can meet most single-phase / three-phase source.
Image 3

• “Direct output” mode, User can set output value by physical key or touch screen.
• Three-phase unit adjustment or single phase adjustment.
Image 4

• “Rotary Knob” mode, User can set in clockwise direction or anticlockwise direction.
Image 5

• Touch “Calibration point” of screen for “% setting”.
Image 6

• “Remote Control Box (option)”, by operation of coarse tuning—fine tuning, can calibrate analog meter.
Image 7

5. Specifications
5.1 Three-Phase Voltage / Current Output

Voltage Range

Resolution

Stability
( %/min )

Accuracy
±(ppm of reading + ppm of range) [1]

Max
Burden (mA)

Class 0.05

Class 0.02

Class 0.05

Class 0.02

57.7 V

0.1 mV

0.01

0.005

300 + 200

120 + 80

250

100 V

1 mV

0.01

0.005

300 + 200

120 + 80

150

220 V

1 mV

0.01

0.005

300 + 200

120 + 80

60

380 V

1 mV

0.01

0.005

300 + 200

120 + 80

40

Note [1]: (ppm = parts per million) (e.g., 10ppm = 0.001%).

Current Range

Resolution

Stability

( %/min )

Accuracy

±(ppm of reading + ppm of range) [1]

Max

Burden (V)

Class 0.05

Class 0.02

Class 0.05

Class 0.02

200 mA

1 μA

0.01

0.005

300 + 200

120 + 80

30

500 mA

1 μA

0.01

0.005

300 + 200

120 + 80

15

1 A

10 μA

0.01

0.005

300 + 200

120 + 80

15

2 A

10 μA

0.01

0.005

300 + 200

120 + 80

4

5 A

10 μA

0.01

0.005

300 + 200

120 + 80

4

20 A

0.1 mA

0.01

0.005

300 + 200

120 + 80

1

• Three-phase voltage output: 1 V~660 V, Degree of distortion:< 0.5%

• Three-phase current output:20 mA~30 A,Degree of distortion:< 0.2%;

• Voltage short circuit, current open-circuit and overload protection

5.2 Frequency / Phase / Harmonic

Symmetry

Voltage superior to 0.2%;Current superior to 0.5%;Phase superior to 0.5°

Frequency

Range:45 Hz~70 Hz;

Adjustment fineness:0.001 Hz;

Accuracy:±0.02 Hz (class 0.05), ±0.01 Hz(class 0.02)

Phase

Range:0.000 0°~359.999 9°;

Adjustment fineness:0.001°;

Accuracy:±0.02°(class 0.05), ±0.01°(class 0.02)

Harmonic

2nd~21st harmonic;

Amplitude 0~25% adjustable;

Phase 0~359.99°adjustable

5.3 Three-Phase Power Output

Current Range

Stability

( %/min )

Accuracy

( ± %*FS)[2]

Class 0.05

Class 0.02

Class 0.05

Class 0.02

Active power |cosφ|≥0.5

0.01

0.005

0.05

0.02

Reactive power |sinφ|≥0.5

0.02

0.01

0.1

0.05

Apparent power

0.02

0.01

0.1

0.05

Power factor

0.02

0.01

0.1

0.05

Note [2] :FS= voltage range× current range

• Power factor setting range:-1.000 000…0.000 000…1.000 000

5.4 Three-Phase Voltage / Current Input

Type

Range

Resolution

Accuracy

±(ppm of reading + ppm of range) [1]

Direct Input

Clamp Current Input [3]

AC Voltage

ACV

57.7 V

0.1 mV

300 + 200

——

100 V

1 mV

300 + 200

——

220 V

1 mV

300 + 200

——

380 V

1 mV

300 + 200

——

AC Current

ACI

1 A

10 μA

300 + 200

——

5 A

10 μA

300 + 200

• Voltage input:6 V~456 V,Current input:0.1 A~6 A
• Frequency:45 Hz~70 Hz,Accuracy:± 0.01 Hz
• Phase:0.000°~359.999°,Accuracy:± 0.02°;

5.5 Three-Phase Power Input

Type

Accuracy

Direct Input

Clamp Current Input

Active power

± 0.05%*FS

± 0.2%*FS

Reactive power

± 0.1%*FS

± 0.5%*FS

Apparent power

± 0.1%*FS

± 0.5%*FS

Power factor

± 0.1%

± 0.5%

5.6 DC Voltage / Current Output

Voltage Range

Resolution

Stability

( %/min )

Accuracy

±(ppm of reading + ppm of range) [1]

Max

Burden (mA)

Class 0.05

Class 0.02

Class 0.05

Class 0.02

75 mV

0.1 μV

0.005

0.005

300 + 200

120 + 80

10

1 V

10 μV

0.005

0.005

300 + 200

120 + 80

10

10 V

0.1 mV

0.005

0.005

300 + 200

120 + 80

10

30 V

0.1 mV

0.005

0.005

300 + 200

120 + 80

500

100 V

1 mV

0.005

0.005

300 + 200

120 + 80

150

300 V

1 mV

0.005

0.005

300 + 200

120 + 80

50

600 V

1 mV

0.005

0.005

300 + 200

120 + 80

25

Current Range

Resolution

Stability

( %/min )

Accuracy

±(ppm of reading + ppm of range) [1]

Max

Burden (V)

Class 0.05

Class 0.02

Class 0.05

Class 0.02

1 mA

10 nA

0.01

0.005

300 + 200

120 + 80

10

5 mA

10 nA

0.01

0.005

300 + 200

120 + 80

10

20 mA

100 nA

0.01

0.005

300 + 200

120 + 80

10

100 mA

1 μA

0.01

0.005

300 + 200

120 + 80

10

1 A

10 μA

0.01

0.005

300 + 200

120 + 80

10

20 A 0.1 mA 0.01 0.005 300 + 200 120 + 80 10

• DC voltage output range:10 mV~660 V,ripple factor:< 1%;
• DC current output range:0.1 mA~22 A,ripple factor:< 1%
• Voltage short circuit, current open-circuit and overload protection

5.7 DC Meter(Transmitter)

Range

Measurement Range

Accuracy

Measurement Range of Ripple

Accuracy of Ripple

1 V

± ( 0~1.2 ) V

± 0.01%*RG

0~30 mV

± 1 mV

10 V

± ( 0~12 ) V

± 0.01%*RG

0~300 mV

± 10 mV

2 mA

± ( 0~2.4) mA

±0.01%*RG

0~60 μA

± 2 μA

20 mA

± ( 0~24 ) mA

± 0.01%*RG

0~600 μA

± 20 μA

• Response time:measurement range:0~1000 ms,accuracy:± 40 ms
5.8 Test Energy Meter(option)

Type

Accuracy

Class 0.05

Class 0.02

Active electrical energy

± 0.1%*reading

± 0.05%* reading

Reactive electrical energy

± 0.2%* reading

± 0.1%* reading

• Energy pulse output:full range for 60 kHz
• Energy pulse input:max frequency is 1 kHz,pulse level:3 V~12 V
• Constant setting of Electrical energy meter:1…1000000 imp./kwh or 1…1000000 imp./ws
5.9 U4 Voltage Output(option)

• Voltage range:100 V, 380 V
• Output range:(0~110)%*RG
• Accuracy:± 0.05%*RG
• Max output power:10 VA
• Frequency range:45 Hz~55 Hz
• The function is used to test Synchrometer.

6. General Specifications

Power Supply

AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz

Temperature Performance

Working temperature:0°C~45°C;

Storage temperature:-20°C~70°C

Humidity Performance

Working humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C

Storage humidity:(20%~80%) R·H,non-condensing

Interface

RS232

7. Ordering Information

 
 